B braun perfusor space коды ошибок

1.0

Perfusor® Space, 1.0 gb

Unit Diagnosis / Calibration

The alarms of the Perfusor® Space are classified in 5 categories.

These categories are listed hereafter according to their

importance.

Alarm advice

In case of unacceptable inputs corresponding messages are

displayed (e.g. «Caution! Rate out of range», «The parameter

cannot be changed») and a beep sounds.

Pre-alarm

Pre-alarms are triggered several minutes (depending on the

service settings) before the operating alarms.

Reminder alarm

A reminder alarm is triggered if the device is not operated for

two minutes when input or operation was not finished.

Operating alarm

In case of an operating alarm the infusion is stopped. An

audible signal is released, the red LED flashes and a staff call

is triggered. The message «Alarm» and the cause of the alarm

appear on the display.

Device alarm

The most important alarms and error codes as well as their

meaning and possible fault clearance are specified in the

following lists.

Note

The device should be checked after every repair or service

«Device Check»

pg. 2 —

9).

2

(see

2 — 3

Дозатор шприцевой Perfusor Space ф.B Braun, Германия

udsasha Дата: Четверг, 20.Янв.2011, 12:14 | Сообщение # 1

Участник

Инженер

OFFлайн

Российская Федерация



Сургут

Ругается на датчик давления ошибка «1022». Как от неё избавиться? Или только в сервис?
.

ЗДЕСЬ по ссылке познавательная информация.

Темы о проблемах со шприцевыми дозаторами (насос шприцевой) серии Perfusor — http://www.medteh.info/forum/59-0-1-0-10-1-[Perfusor]

 
STaras Дата: Четверг, 20.Янв.2011, 18:53 | Сообщение # 2

инженер-электроакустика, ультразвук

OFFлайн

Украина



Ровно

Может 2022?

Quote (udsasha)

Или только в сервис?

— или посмотреть, чего ж там в датчике давления сломалось, и попытаться починить

Quote (udsasha)

Или только в сервис


Чти вело свое – ибо не ты на нём едешь, но оно везёт тебя

 
Andrey0612 Дата: Четверг, 03.Фев.2011, 20:01 | Сообщение # 3

Участник

инженер

OFFлайн

Российская Федерация



Ижевск

Судя по таблице кодов ошибок «1022 FUP pressure zero test fail» —> Написано, провести калибровку..
По моему только в сервис unknown

Сообщение отредактировал Andrey0612Четверг, 03.Фев.2011, 20:03

 
udsasha Дата: Пятница, 04.Фев.2011, 08:41 | Сообщение # 4

Участник

Инженер

OFFлайн

Российская Федерация



Сургут

Отписал в сервис, а там как начальству будет угодно.

 
feniks71 Дата: Среда, 01.Фев.2012, 11:03 | Сообщение # 5

Заглянувший

Мед. аппараты и системы

OFFлайн

Российская Федерация



Москва

1015 ошибка Perfusor Space B.Braun, поменял поршень с движком и схемку управления лапками держалки шприц поршня подсоединил все и вот такое теперь. все контакты и соединения проверил по 3 раза. Что за ошибка кто знает?

Сообщение отредактировал feniks71Среда, 01.Фев.2012, 11:06

 
Alexium Дата: Среда, 01.Фев.2012, 13:00 | Сообщение # 6

Инженер медтехник

OFFлайн

Российская Федерация



Саратов

Проверьте пищалку (динамик)

 
feniks71 Дата: Среда, 01.Фев.2012, 14:36 | Сообщение # 7

Заглянувший

Мед. аппараты и системы

OFFлайн

Российская Федерация



Москва

Quote (Alexium)

Проверьте пищалку (динамик)
Прикрепления: 5750984.jpg(10Kb)

Мелкий динамик высокой частоты пищит при ошибке громко и ясно…. Сейчас разобрал все проверил =/ Спасибо, динамик большой сломан.

Сообщение отредактировал feniks71Среда, 01.Фев.2012, 14:41

 
Texnikims Дата: Понедельник, 12.Мар.2012, 17:33 | Сообщение # 8

Техник

Сервисный Инженер

OFFлайн

Российская Федерация



МО Люберцы

Уважаемые коллеги!
Помогите разобраться с насосом (помпой, перфузором) BBraun Perfusor Spase. Проблема в том, что он постоянно требует *заменить шприц*,ёникаких ошибок не выдаёт кроме *обслужить аккумулятор*, но я его от сети гоняю … Вставляю шприц (типа поменял) нажимаю ОК и ничего не происходит, механизм двигающий шприц отведён в право до упора и абсолютно не двигается, фиксаторы поршня шприца в закрытом состоянии! После нажатия клавиши ОК раздаётся звук внутри корпуса (включается мотор на секунду, но механизм не двигается). В инструкции ничего интересного нет! Разобрать аппарат не получается, нет ни одного винта! Забыл сказать, что аппарат проходит самодиагностику и ни на что не ругается!

Сообщение отредактировал TexnikimsПонедельник, 12.Мар.2012, 17:37

 
Alexium Дата: Вторник, 13.Мар.2012, 19:49 | Сообщение # 9

Инженер медтехник

OFFлайн

Российская Федерация



Саратов

Скачайте сервис мануал — http://www.medteh.info/load/41-1-0-1652

 
ax6842 Дата: Вторник, 16.Окт.2012, 19:47 | Сообщение # 10

Заглянувший

эл.мех.

OFFлайн

Российская Федерация



Барнаул

Quote (Texnikims)

Уважаемые коллеги! помогите разобраться с насосом(помпой,перфузором) BBraun Perfusor Spase. Проблема в том что он постоянно требует *заменить шприц*,никаких ошибок не выдает кроме *обслужить аккамулятор* но я его от сети гоняю… вставляю шприц (типа поменял) нажимаю ок и ничего не происходит,механизм двигающий шприц отведен в право до упора и абсолютно не двигается,фиксаторы поршня шприца в закрытом состоянии! после нажатия клавиши ок раздается звук внутри корпуса (включается мотор на секунду но механизм не двигается) в инструкции ничего интересного нет! разобрать аппарат не получается,нет ни одного винта! забыл сказать что аппарат проходит самодиагностику и не на что не ругается!

Та же самая проблема:
1. Двигается зажим, но шприц не зажимает (выдвигается полностью, при нажатие двигается вперед, назад, пытается захватить) и просит заменить шприц.
2. При выключении задвижка остается на месте. Хотя по инструкции должна возвращаться в исходное состояние (втянут).
3. С главного меню даже немогу в настройки инфузии войти. Прогроматора нет.. Самоконтроль проходит

Скачал ман предложенный Alexium, надеюсь поможет.
Но интересует, есть ли сброс настроек на «заводские» (Заранее извиняюсь если глупые вопросы, данные приборы вижу впервые), дабы получить доступ к меню настройки подачи, объема шприца.

 
ax6842 Дата: Среда, 17.Окт.2012, 15:17 | Сообщение # 11

Заглянувший

эл.мех.

OFFлайн

Российская Федерация



Барнаул

проблема решена, путем обмана прибора:
Вставил шприц, нажал на головке привода датчик, в момент закрытия держателей.
С 3-его раза прибор вывел список шприцов и заработал как положено.

 
Quex Дата: Четверг, 26.Сен.2013, 16:20 | Сообщение # 12

Участник

техник

OFFлайн

Российская Федерация



Ч

При включении самодиагностика не запускается, ошибок не выдает, а начинает мигать всеми своими тремя светодиодами и пиликать. На экране ничего нет. Уже два таких. Что-то можно сделать?

 
alex219 Дата: Суббота, 28.Сен.2013, 01:13 | Сообщение # 13

Участник

инженер по обслуживанию медтехники

OFFлайн

Российская Федерация



Хабаровск

проверьте шлейф от блока к панели индикатора, выпадал от «бережного обращения»

 
alex219 Дата: Суббота, 28.Сен.2013, 01:31 | Сообщение # 14

Участник

инженер по обслуживанию медтехники

OFFлайн

Российская Федерация



Хабаровск

Добрый день всем. Принесли в ремонт после хорошего падения.Тест проходил просил провести замену шприца , при этом каретка выдвинута полностью подвижные губки захвата стоят на месте(при нажатии на сенсор в центре или установки шприца). В результате выяснилось что от «бережного обращения» отломилась ось потенциометра в механизме захвата, заклеил клеем пвх для лодок, держится.В итоге при тестировании дает теперь ошибку 1203 в мануале пишут о необходимости проведения каллибровки при помощи сервисной программы, губки захвата шприца пытаются произвести захват а каретка стоит колом.В описании процесса пользования сервисной программой не нашел как каллибруется шаговый двигатель.Самой программы к сожелению нет.Фото прилагаю.Если есть программа готов расмотреть вариант обмена.с уважением александр.

 
alex219 Дата: Суббота, 28.Сен.2013, 01:39 | Сообщение # 15

Участник

инженер по обслуживанию медтехники

OFFлайн

Российская Федерация



Хабаровск

фото прилагаю

 
kurvla79 Дата: Среда, 16.Окт.2013, 10:38 | Сообщение # 16

Участник

электромеханик

OFFлайн

Российская Федерация



там

Привет Всем!
Очередная машинка написала ошибку 1119 ….
Я понимаю, что это- LC display defective… у буржуев лечится это просто — Exchange operating unit, но может есть и «наш» путь?

 
kurvla79 Дата: Четверг, 17.Окт.2013, 11:59 | Сообщение # 17

Участник

электромеханик

OFFлайн

Российская Федерация



там

«Свой» путь есть!!!)))
Может кому пригодится: backlight там сделан светодиодный — 5 белых светодиодов последовательно…
Анод этой линейки приходит на 8ю ногу шлейфа LCD, катод — на 9-ю! …. и далее на шлейф на основной блок- 7 дорожка-катод, 8я — анод… и ошибки больше нет!!! drinks

Сообщение отредактировал kurvla79Пятница, 18.Окт.2013, 10:24

 
voldyr72 Дата: Среда, 12.Фев.2014, 07:06 | Сообщение # 18

Стажер

сервисный инженер

OFFлайн

Российская Федерация



Иркутск

Здравствуйте коллеги!Подскажите пожалуйста,на инфузомате B.Braun spase при включении моргает аларм 2013.Что это?Сервисной инструкции,к сожалению нет.Заранее спасибо.

 
kurvla79 Дата: Среда, 12.Фев.2014, 21:03 | Сообщение # 19

Участник

электромеханик

OFFлайн

Российская Федерация



там

internal error

 
kurvla79 Дата: Пятница, 14.Фев.2014, 07:46 | Сообщение # 20

Участник

электромеханик

OFFлайн

Российская Федерация



там

…Случилось тоже самое только на перфузоре B.Braun spase (номер ошибки 1013 (у перфузоров свой ряд)- тоже внутренняя ошибка) — оставил прибор включенным до полного разряда… При заряде ошибки уже не было..

 
lin72 Дата: Понедельник, 17.Мар.2014, 08:31 | Сообщение # 21

Заглянувший

инженер

OFFлайн

Российская Федерация



Тюмень

Здравствуйте коллеги! Подскажите пожалуйста, BBraun Perfusor Space при включении выдает ошибку 1121. Что это, возможно ли исправить?

 
kurvla79 Дата: Вторник, 18.Мар.2014, 08:32 | Сообщение # 22

Участник

электромеханик

OFFлайн

Российская Федерация



там

по мануалу это red led on defect….,т.е. ошибка красного светодиода.. при включении аппарат проводит тест панели управления… ошибка из той же серии что и 1119 (смотрите выше) … скорее всего аппарат залили и потеряна связь со светодиодом(дорожки в панели очень тонкие — не выдерживают русского обращения))..

Сообщение отредактировал kurvla79Вторник, 18.Мар.2014, 08:58

 
sqvair Дата: Среда, 23.Апр.2014, 14:23 | Сообщение # 23

Стажер

Инженер

OFFлайн

Российская Федерация



Санкт-Петербург

Для Space. Все тем, у кого постоянно требует провести замену шприца. На головке поршня снимаете резиновую насадку. Под ней увидите кнопку заподлицо с корпусом. Нажимаете её карандашом или шариковой ручкой и не отпуская кнопку пальчиками двигаете захваты шприца (несколько раз открыть-закрыть). После чего отпускаете кнопку. Должно помочь.

Цитата lin72 ()

Подскажите пожалуйста, BBraun Perfusor Space при включении выдает ошибку 1121. Что это, возможно ли исправить?

Проверьте шлейф панельки. (Замена панели)

 
Post_digg Дата: Суббота, 24.Май.2014, 11:20 | Сообщение # 24

Стажер

Сервис-инженеррадиотехник

OFFлайн

Российская Федерация



Иркутск

Доброго времени суток, коллеги! Perfusor space-вский Би-браун выдает ошибку 1216 «overvoltage test fail; Voltage transformer defective; Exchange processor PCB;» тест перенапряжения терпит неудачу; дефектный трансформатор Напряжения; Поменять процессор PCB. Может кто поделиться мыслями? Внешний осмотр платы произвел, основные резисторы-диоды прозвонил и ошибки не выявил… Подскажите пожалуйста, куда копать?

 
kurvla79 Дата: Суббота, 24.Май.2014, 21:35 | Сообщение # 25

Участник

электромеханик

OFFлайн

Российская Федерация



там

в сервиснике есть две таблицы UTS (unit test step(таблица тестов)) и TS (troubleshooting(поиск неисправностей)).. по ним попробуйте отследить свою ошибку — Exchange processor PCB… там есть описание симптомов и две колонки:если «да», то…И если «нет»,то….

 
Post_digg Дата: Воскресенье, 25.Май.2014, 10:24 | Сообщение # 26

Стажер

Сервис-инженеррадиотехник

OFFлайн

Российская Федерация



Иркутск

Ой-е! А я и проглядел))) Да простят коллеги невнимательность мою) Спасибо за подсказку)

 
bedmed Дата: Воскресенье, 25.Май.2014, 16:31 | Сообщение # 27

Стажер

Инжинер по м/о

OFFлайн

Российская Федерация



Иркутск

[s]День добрый, коллеги.
По моей статистике:
1. Роняют на пол (ломаются датчики, гнётся винт привода, на некоторых моделях отлетают наклейки со штрихами);
2. Роняют при дезинфекции в ведро с раствором (молчат!!! аккумулятор даёт питание, выходят из строя ленточные контакты кнопок управления или всё полностью окисляется);
3. Из-за перебоев питания (аккумуляторы (замена, или только от сети (вытащить батарею)), выход из строя платы индикации, выход из строя контролера с прошивкой).
1 механика — для этого нужно выправить винт и пройтись леркой (в месте загиба может показывать «инфузию»), датчики перепаиваются комплектом, можно на барабане маркером нанести штрихи нужной толщены или вырезать диск из поликарбоната (бутылки из под газ. воды) и маркером разметить.
2 зачищаю контакты (до целого нормального сечения) покупаете «клей для восстановления токоведущих частей», дальше по инструкции.
3 С рабочего аппарата сливаете прогу, покупаете контроллер и заливаете.

 
Post_digg Дата: Понедельник, 26.Май.2014, 11:08 | Сообщение # 28

Стажер

Сервис-инженеррадиотехник

OFFлайн

Российская Федерация



Иркутск

Кст, по поводу прошивок: А что вы скажете по поводу программатора на ЛПТ-порте? Простейший шлейф из 5 проводов напрямую на контроллер? С учетом того что комп без земли. Какие подводные камни могу встретить в данном случае с применением ПониПрог?

 
kurvla79 Дата: Вторник, 27.Май.2014, 19:27 | Сообщение # 29

Участник

электромеханик

OFFлайн

Российская Федерация



там

в сервиснике есть две таблицы UTS (unit test step(таблица тестов)) и TS (troubleshooting(поиск неисправностей)).. по ним попробуйте отследить свою ошибку — Exchange processor PCB… там есть описание симптомов и две колонки:если «да», то…И если «нет»,то….

 
san1987 Дата: Вторник, 12.Авг.2014, 09:30 | Сообщение # 30

Стажер

инженер

OFFлайн

Российская Федерация



ЯНАО

Здравствуйте!
У меня такая проблема со шприцом: не двигается задвижка (не задвигается и не выдвигается), лапки не двигаются (двигатель работает на раздвижение, но при нажатии на кнопку-датчик они не смыкаются).

 
san1987 Дата: Четверг, 14.Авг.2014, 14:33 | Сообщение # 31

Стажер

инженер

OFFлайн

Российская Федерация



ЯНАО

Подскажите, какой потенциометр стоит в контроле положения захватов поршня шприца.

 
MAPPER_1222 Дата: Пятница, 17.Июл.2015, 11:38 | Сообщение # 32

Техник

Инженер и этим всё сказано…

OFFлайн

Российская Федерация



Москва

Кто-нибудь сталкивался на инфузомате ошибка «2014» постоянно при включении. Динамик поменял всё тоже самое.

 
Alhimik11 Дата: Понедельник, 31.Авг.2015, 15:16 | Сообщение # 33

Участник

Медтехник

OFFлайн

Российская Федерация



Архангельск

Что-то не понятно, как исправить «1119» ошибку blush На вид всё чисто!

 
MAPPER_1222 Дата: Среда, 02.Сен.2015, 09:31 | Сообщение # 34

Техник

Инженер и этим всё сказано…

OFFлайн

Российская Федерация



Москва

Ошибка «2111» при выключении зависает мигает красным диодом со звуком.
Кто-нибудь сталкивался?

 
MAPPER_1222 Дата: Среда, 02.Сен.2015, 09:34 | Сообщение # 35

Техник

Инженер и этим всё сказано…

OFFлайн

Российская Федерация



Москва

Цитата Alhimik11 ()

Что-то не понятно, как исправить 1119 ошибку blush На вид все чисто!

Если есть возможность подкиньте другой дисплей. Возможно пропадет

 
Alhimik11 Дата: Вторник, 22.Сен.2015, 09:11 | Сообщение # 36

Участник

Медтехник

OFFлайн

Российская Федерация



Архангельск

Замена дисплея проблему не решила. Принесли ещё один такой аппарат, выдаёт ошибку «1212»!
Может кто сталкивался?! blush

 
vdv777 Дата: Среда, 28.Окт.2015, 22:06 | Сообщение # 37

Стажер

инженер

OFFлайн

Российская Федерация



Москва

При ошибке 1119 — помогает замена лицевой панели с клавиатурой, как правило проблема в заломе плоского шлейфа или в окислении/отгнивании контактов внутри передней панели (когда зальют какой-нибудь гадостью), редко бывает, что просто выскакивает шлейф из разъёма внутри аппарата.
1112 скорее всего тоже — замена лицевой клавиатурной панели. (Когда нормальная связь с клавиатурной панелью нарушается выскакивают подобные ошибки).
P.S. сам дисплей при этом как правило остаётся исправным.

А вот сам сейчас столкнулся на ряде Спейсов с такой проблемой — после установки шприца просит второй раз произвести смену шприца, жмёшь кнопку ОК, снова прижимает шприц металлической выдвижной пластиной и со второго раза шприц определяется, появляется выбор моделей шприцов и дальше всё работает как надо. Но такая ерунда каждый раз при установке нового шприца. Может кто уже победил сей глюк? (Думается что требует калибровки.)

 
Андрей_Алтай Дата: Вторник, 15.Дек.2015, 02:37 | Сообщение # 38

Стажер

набрали по объявлению

OFFлайн

Российская Федерация



Барнаул

При ошибке 1119 помогла замена шлейфа. Труднее было его подобрать для замены. 30pin, шаг 1мм, длина ~100мм.

 
tuman_777 Дата: Пятница, 18.Дек.2015, 23:53 | Сообщение # 39

Участник

электронщик

OFFлайн

Российская Федерация



Российская Федерация

Цитата vdv777 ()

А вот сам сейчас столкнулся на ряде Спейсов с такой проблемой — после установки шприца просит второй раз произвести смену шприца, жмёшь кнопку ОК, снова прижимает шприц металлической выдвижной пластиной и со второго раза шприц определяется, появляется выбор моделей шприцов и дальше всё работает как надо. Но такая ерунда каждый раз при установке нового шприца. Может кто уже победил сей глюк? (Думается что требует калибровки.)

Шприцы, скорее всего, не из списка.

 
V_A_L Дата: Четверг, 24.Дек.2015, 07:06 | Сообщение # 40

Участник

инженер

OFFлайн

Российская Федерация



Томск

Коллеги!
Кто-нибудь по нищете заменял Ni-MH аккумуляторы для Спейсов в Battery pack SP (pn 0871 3180)?
Даже при установке свежих, заряженных элементов перфузор считает, что аккумулятор не заряжен, но на «Обслуживание батареи» не выходит.
Подозреваю, что нужно сбросить счетчик в контроллере батареи. Как?

 

Perfusor® Space

Service Manual

Version 1.3 English

0

This Service Manual is valid for:

Designation

Part No.

Infusion syringe pump Perfusor® Space . . . . . . . . . . .

. 0871 3030

This Service Manual is available under

Designation

Part No.

the following part number:

Service Manual Perfusor® Space, English. . . . . . . . . . .

8713 9020

Languages of this Manual

The Service Manual for this unit can be supplied in the following

languages:

Designation

Part No.

Service Manual Perfusor® Space, German . . . . . . . . . .

8713 9010

Service Manual Perfusor® Space, English (US). . . . . .

8713 9020U

Service Manual Perfusor® Space, French . . . . . . . . . . .

8713 9030

2

Perfusor® Space, 1.2 gb

Table of Contents 0

Important Preliminary Remarks

Service Work

Page

0 — 5

Technical Safety Checks

Page

0 — 5

Current Versions

Page

0 — 5

Revision Service

Page

0 — 5

Quality Management

Page

0 — 6

Checks and Repair

Page

0 — 6

Notes on ESD

Page

0 — 6

Spare Parts and Test Equipment

Page

0 — 7

Setting Off

Page

0 — 7

Contact Persons

Technical Training

Page

0 — 11

Entry for Technical Training

Page

0 — 11

Ordering of Spare Parts and Test Equipment

Page

0 — 11

Service Hotline International

Page

0 — 11

Return of Spare Parts and Test Equipment

Page

0 — 11

Safety Officer

(§ 30 MPG)

Page

0 — 11

Translation

Page

0 — 11

System Overview

Description

Page

1 — 1

System Overview

Page

1 — 1

Physical Construction

Page

1 — 2

Function

Page

1 — 3

Unit Software

Page

1 — 6

Service Program

Page

1 — 7

Technical Data

Page

1 — 12

Options

Page

1 — 12

Accessories

Page

1 — 12

Unit Diagnosis / Calibration

General

Page

2 — 1

Alarms and Error Codes

Page

2 — 3

The Most Important Error Modes

Page

2 — 8

Device Check

Page

2 — 9

Calibration

Page

2 — 14

Procedural Instructions for Calibration

Page

2 — 14

Trouble Shooting

Page

2 — 29

Disassembly / Assembly

General

Page

3 — 1

Battery Module

Page

3 — 9

Unit Foot

Page

3 — 11

Operating Unit

Page

3 — 12

Upper Part of Housing

Page

3 — 17

Release Button

Page

3 — 19

Loudspeaker

Page

3 — 19

Drive

Page

3 — 20

Syringe Holder with Piston Brake

Page

3 — 27

Processor PCB

Page

3 — 34

Assembly / Installation

Page

3 — 35

Checks after Repair

Page

3 — 47

Perfusor® Space, 1.0 gb

0 — 3

Servicing the Unit

Cleaning

Page

4

— 1

Servicing the Battery

Page

4

— 1

Technical Safety Check (TSC)

Perfusor® Space 1

Technical Safety Check (TSC)

Power Supply SP 1

Procedural Instructions on the TSC

Visual Inspection

Page

7

— 1

Electrical Safety

according to IEC/EN 60601-1

or VDE 0750 and VDE 0751

Page

7

— 2

Functional Inspection Perfusor® Space

Page

7

— 3

Functional Inspection Power Supply SP

Page

7

— 6

Test Equipment and Special Tools

Test equipment

Page

8

— 1

Special Tools

Page

8

— 3

Spare Parts List

Page

9

— 1

Revision Documentation

Description of Version

Page

10

— 1

Version List of the Individual Pages

Page

10

— 1

Index

Page

11

— 1

0 — 4

Perfusor® Space, 1.0 gb

Important Preliminary Remarks 0

Service Work

The present manual is for your information only. The possession of

this manual does not authorize the performance of service work.

Service tasks may only be executed by persons, who

— have received appropriate training on the system from

B. Braun

— are included in the revision service

— possess the necessary test equipment and mechanical aids,

and

— fulfill the personal requirements (training and knowledge).

Technical Safety Checks

The user is obliged to perform or to have performed the Technical

Safety Checks on those medial products for which these checks

have been prescribed by the manufacturer and to carry them out

according to the indications of the manufacturer as well as the

generally approved technical standards while adhering to the

periods stated (§ 6 MP BetreibV).

B. Braun also recommends training on the Technical Safety

Checks, or to perform at least the steps indicated in the current

version of the manual, as:

— the TSC requires that the instructions in the manuals are

observed

— the manuals are a reference for measurements

— depending on the unit type, the Service Program must be

called which may lead to a dangerous unit condition in case

of inappropriate operation. Furthermore, a special service

connector may be necessary.

Current Versions

This manual version corresponds to the state when the manual

was written. B Braun reserves the right to make technical

modifications. The state of the revision is indicated by the index

number in the footer of every page.

Revision Service

The possession of this manual does not automatically mean

inclusion in the revision service. You will be included in the

revision service after:

— technical training by B. Braun Melsungen or

— a written order placed with the sales department of B. Braun

(fee required).

Perfusor® Space, 1.1 gb

0 — 5

0

Important Preliminary Remarks

Responsibility of the Manufacturer

The manufacturer, person who assembles, installs or imports the

device can only be held responsible for safety, reliability and

performance if

— mounting, enhancements, new settings, changes or repairs

are carried out by duly authorized persons,

— the electrical installation in the corresponding room meets

the requirements of the VDE 0107, VDE 0100 part 710 or

IEC 60364-7-710 and the national standards,

— the device is used in accordance with the instructions for use

and the Service Manual,

— the Technical Safety Checks are performed at regular

intervals,

— a current manual which corresponds to the revision state is

used when carrying out maintenance, repair and service,

— the service technician takes part in the revision service,

— the technician has participated in a technical training course

for the specific B. Braun unit.

Quality Management

B. Braun is certified in accordance with DIN EN ISO 9001 and

ISO 13485. This certification also includes maintenance and

service.

The unit has the CE label. The CE label confirms that the device

corresponds to the “Directive of the Council for Medical Products

93/42/EC” of June 14, 1993.

Checks and Repair

Training may only be performed by B. Braun. The possession of the

manual does not authorize the performance of repairs. The

instructions on electrostatic sensitive components (ESD

standards) must be observed.

After repair a device check or diagnosis is to be carried out.

Notes on ESD

Semiconductors can be destroyed by electrostatic discharge.

Especially MOS components can be damaged by interference from

electrostatic fields, even without discharge via contact. This type

of damage is not immediately recognizable. Unit malfunctions

can even occur after a longer period of operation.

0 — 6

Perfusor® Space, 1.1 gb

Important Preliminary Remarks

0

Fig.: 0 — 1

Spare Parts and Test Equipment

Setting Off

Each workstation must be equipped according to the recommendations with the necessary static protective measures, if ESD components or boards are handled.

Each workstation must be equipped with a conductive table surface. The conductive surface, the soldering iron or the soldering stations must be grounded via protective resistors.

Chairs must be of antistatic design. The floor or floor mats should be of electrically conductive material.

Personnel must wear conductive wristbands which are connected to a central ground potential via protective resistors, e.g. the ground contact of a wall outlet. Furthermore it is recommended that personnel wear cotton clothing and electrically conductive shoes to prevent electrostatic charge.

Only use original spare parts from the manufacturer. Do not tamper with assembly groups which can only be exchanged completely. The spare parts required are listed in the repair descriptions.

Service personnel are responsible for the calibration of their test equipment. Original test equipment can be calibrated at the works of B. Braun. Further information is available upon request.

Additional notes and warnings are set off as follows:

Note

Is used for additional or special notes concerning information and working steps.

CAUTION

Is used for working steps which may result in damage to the unit, system or to a connected device.

WARNING

IS USED FOR WORKING STEPS WHICH MAY RESULT IN PERSONAL INJURY.

Perfusor® Space, 1.1 gb

0 — 7

0

Important Preliminary Remarks

References to chapters are shown as follows (see “Setting Off“ pg. 0 — 7)

References to figures and tables are shown as follows

Fig.: 2 — 3 or Table 2 — 1

References to item numbers in figures are shown as follows (Fig.: 1 — 1 / Item 1)

In this case “Fig.: 1 — 1“ is the figure number and “Item 1“ the item number within the figure.

When the Service Manual is stored as pdf-file, these references are displayed green. Click with the mouse button on a reference to jump to the corresponding source.

Menu commands are described as: Menu File.

0 — 8

Perfusor® Space, 1.1 gb

Important Preliminary Remarks

0

List of Abbreviations

Abbreviations which are not generally known, but are used in this

manual, are listed below.

CAN

Controller Area Network

CE

Communauté Européenne

(European Communities)

CS

Calibration Step

DIN

Deutsche Industrie Norm

(German Industrial Standard)

EN

European Standard

ESD

Electrostatic Discharge

FuP

Function Microprocessor

IEC

International Electrotechnical

Commission

ISO

International Standardization

Organization

ISP

Infusomat® Space

ISPS

Infusomat® Space, Silicon

ISPP

Infusomat® Space, PVC

KuP

Monitoring Microprocessor

LCD

Liquid Crystal Display

MOS

Short for the following

company name:

MOS Technology, Inc.

(Commodore Semiconductor

Group)

PCA

Patient-Controlled Analgesia

PSP

Perfusor® Space

SP

Space (System)

SPC

SpaceCover

SPCC

SpaceCover comfort

SPCS

SpaceCover standard

SPCO

SpaceCom

SPCT

SpaceControl

SPS

SpaceStation

TEMP

Temperature

TS

Troubleshooting Step

TSC

Technical Safety Checks

Perfusor® Space, 1.2 gb

0 — 9

0

Important Preliminary Remarks

UTS

Unit Test Step

VDE

Verband der Elektrotechnik,

Elektronik und

Informationstechnik e.V.

(German electrical engineering

association)

0 — 10

Perfusor® Space, 1.2 gb

Contact Persons 0

Technical Training

Via local representative.

Entry for Technical Training

Application for a technical training course must be made via the

responsible representative.

Ordering of Spare Parts and Test Equipment

Please contact your local B. Braun subsidary.

International Technicians (Intercompany)

Nadja Machal

Fax:

+49 5661 / 75 — 47 89

e-mail:

nadja.machal@bbraun.com

Service Hotline International

Karl Tippel, Tanja Kördel

Fax:

+49 5661 / 71 — 35 26

e-mail:

karl.tippel@bbraun.com

e-mail:

tanja.koerdel@bbraun.com

Return of Spare Parts and Test Equipment

B. Braun Melsungen AG

Schwarzenberger Weg 73-79

Wareneingang Werk C

34 212 Melsungen

Germany

Safety Officer

Dr. Ludwig Schütz

(§ 30 MPG)

e-mail: ludwig.schuetz@bbraun.com

Translation

Cs2 Informatik GmbH & Co. KG, Germany

Perfusor® Space, 1.3 gb

0 — 11

0 — 12

Perfusor® Space, 1.0 gb

System Overview 1

Description

The Perfusor® Space (PSP) is according to IEC/EN 60601 resp. IEC/

EN 60601-2-24 a transportable infusion syringe pump for

administrating fluids in the nutritional therapy and infusion

technique as well as for home care applications.

The medical specialist must decide on suitability for application

on the basis of the warranted properties and the technical data.

System Overview

The Space system is a modular design of modern infusion

technology for stationary, mobile or private use. The key modules

and their connection to the peripheral devices are shown in

Fig.: 1 — 1.

All the pump types, Perfusor® Space, Infusomat® Space and

Infusomat® Space P, as well as the other devices of the system are

1

of modular design. Up to three pumps can be connected together

mechanically using L rails on the bottom of the unit and grooves

2

on the top. They can then be fastened to a drip stand or

appropriate rail using the pole clamp.

The SpaceControl module can be used to extend operation. One

3

single pump can be inserted onto this module. The pump is then

connected via connectors to the module.

The SpaceStation module allows the set-up of a complete pump

system with up to 24 pumps. Up to four pumps can be installed in

4

every SpaceStation. The pumps are supplied with power via the

5

integrated power supply and the built-in connectors. The pumps

are connected to the optional SpaceCom via these connectors.

SpaceControl can also be integrated into the system.

Up to six SpaceStations can be set-up as a column with a total of

24 pumps. SpaceStation placed next to each other can be

Fig.: 1 — 1 Space system

connected via special connection cables, if the maximum number

of 24 pumps in maximum three columns is not exceeded.

Legend of fig. 1 — 1:

SpaceCover Standard or SpaceCover Comfort forms the top of

ItemDesignation

each column. Alarms are signalled by a row of LEDs and a

1 SpaceCover

loudspeaker in the SpaceCover Comfort.

2Infusion pump Infusomat® Space

3Infusion syringe pump Perfusor® Space

4SpaceControl

5SpaceStation

Perfusor® Space, 1.0 gb

1 — 1

Physical Construction

5

4

3

6

Fig.: 1 — 2 Perfusor® Space

Legend of fig. 1 — 2:

ItemDesignation

1

Perfusor® Space

6

Connector “P2“ for SpaceStation module, external 12 V DC

2

Drive head

and accessories

3

Syringe holder with piston brake

7 Connector “P3“, connection to SpaceControl module

4

Operating Unit

8

Battery compartment cover

5

Syringe area

1 — 2

Perfusor® Space, 1.0 gb

The Perfusor® Space housing mainly consists of the bottom part

and the upper part.

The battery module is inserted in the rear of the housing upper

part. The opening is covered by the battery compartment cover.

The operating unit is attached to the front of the bottom part with

two hinges. This operating unit covers the area for the syringes.

The complete drive assembly, consisting of lead screw and drive

head with driving tube is located directly behind the syringe area

in the bottom part of the housing. The housing bushing for the

driving tube is located in the side of the housing.

The syringe holder is mounted in the right side of the housing

bottom part.

The processor PCB with the permanently connected external

connectors “P2” and “P3” is located at the bottom of the housing

bottom part.

Function

There are two power options for the Perfusor® Space:

— via the inserted battery module

— via an external 12 V DC power supply (e.g. SpaceStation,

SpaceControl, an external power supply or from an

ambulance car) connected to connector “P2”

The voltage supplied is converted to the internal voltages required

through a voltage transforming and monitoring circuit on the

processor PCB.

An independent circuit in the battery module monitors the battery

cells and controls their charge condition.

The Perfusor® Space is connected to a SpaceControl by connector

“P3”.

The function processor controls all the functions of the Perfusor®

Space. Data is stored in a non-volatile memory which also

controls the external data transfer.

The control microprocessor monitors all important responses of

the function processor to incoming information. If a response

does not correspond with that expected by the control

microprocessor, an error message is generated and the device is

switched to a safe stop state.

The drive motor is monitored by a detector for speed and direction

of rotation. The extended end position of the drive head is

detected by a switch on the processor PCB.

Perfusor® Space, 1.0 gb

1 — 3

The pressure in the infusion system is measured through a strain gauge measuring in the drive head and monitored in the device electronics. The data from the strain gauge is continuously compared with the limit values which are calculated dependent on the selected syringe type and the pressure settings. When the limit values are exceeded an alarm is automatically triggered and the pressure in the infusion system is reduced. The maximum pressure is additionally limited by a second, independent system. This maximum pressure limitation is performed using the motor current control.

The syringe size detection is performed via the syringe holder. The syringe holder is connected to a potentiometer. The syringe size is determined from the resistance of the potentiometer.

The syringe is fixed with the syringe holder and the axial fastening device. The syringe piston is fastened with two claws in the drive head. When a syringe is inserted the syringe piston is held by the piston brake, until the piston has been caught by the claws.

Keyboard and display as well as the syringe area are illuminated.

1 — 4

Perfusor® Space, 1.0 gb

B Braun Perfusor Space User manual

System Overview

1

Fig.: 1 — 3 Block diagram Perfusor® Space

Perfusor® Space, 1.0 gb

1 — 5

Unit Software

Approved Software Versions

688A030032

Basic software

Position

1

2

3

4

5

6

7

8

9

10

Digit

6

8

8

C

0

3

0

0

0

1

688A030035

Improved functions

688A030040

Revision level

Improved functions

Hardware

Software group

Languages French and Swedish added

688B030002

Device type: Perfusor® Space

Fig.: 1 — 4

Improved functions

688B030003

CAN bus functioning

688C030001

Dose calculation

Changed CAN log

688D030001

— Drug list data base

Changed user language

688E030003

Improved functions

Piggyback

Soft limits

688F030006

PCA

Changed claw configuration

Optimized alarm handling

688G030002

Improved functions

1 — 6

Perfusor® Space, 1.0 gb

Service Program

Software Update of the Unit

The instructions for updating the software are supplied with the software itself.

CAUTION

If the device is disconnected while the software is being updated or the device or PC is switched off, a component of the software may be seriously damaged so that repairs are no longer possible. In such a case the software cannot be updated via the PC and the device must be returned to B. Braun.

Approved Version

Note

Please note that text and / or functions of the Service Program may change depending on the software version. The following screen illustrations are only examples and represent the state when the manual was printed.

0.0.28

1.0.0

1.1.2

1.1.3

1.1.4

1.2.1

1.3.5

1.5.0

2.0.1

3.1.0

4.0.0

5.1.0

Starting the Service Program

Note

Installation and further operation of the Service Program is described in its separate instructions for use.

Perfusor® Space, 1.0 gb

1 — 7

Fig.: 1 — 5

Fig.: 1 — 6

Fig.: 1 — 7

1.Start the “HiBaSeD.exe” program (History, Barcode, Service, Drug list) on the PC. The Service Program is loaded and started and the initial window of the Service Program is displayed.

2.Read the notes carefully.

3.Mark the field “I accept all conditions” and then the field “Yes” to confirm that you have read the notes.

Note

Click the field “English” to switch the language of the notes over to English.

4.Enter the password and confirm it by clicking the field “Start”.

The Service Program checks the PC interfaces for connected devices of the Space system. Units that were found are displayed for a short moment on the screen.

1 — 8

Perfusor® Space, 1.0 gb

The work window of the Service Program appears on the screen. All devices recognized are listed in the left column.

Fig.: 1 — 8

5. Activate the desired device from the list on the left in the work window with a double-click. The device data is then displayed below the device name.

Perfusor® Space, 1.0 gb

1 — 9

If the unit software version is not compatible with the Service Program version, a window opens prompting the operator to change the Service Program version. This window displays a compatibility list of the Service Programand unit software versions.

If Service Programand unit software versions are compatible, all the Service Program functions are activated.

Fig.: 1 — 9

Fig.: 1 — 10

1 — 10

Perfusor® Space, 1.0 gb

Fig.: 1 — 11

Fig.: 1 — 12

Service Program Version

1.Open the “HiBaSeD“ window via Help Info …. The current version of the Service Program is shown in this window.

2.Close the window by clicking “OK”.

Compatibility List

1.Open the “Unit — Compatibility” window via Help Compatibility. This window displays the compatibility of the HiBaSeD-version and the unit software version.

2.Close the window by clicking “OK”.

Quit the Service Program

1.Exit the Service Program via Application Quit.

2.Disconnect a power supply which might be connected from the unit.

3.Switch off the unit.

4.Remove the battery module.

5.The device can be restarted after appr. 10 seconds.

Perfusor® Space, 1.0 gb

1 — 11

Technical Data

Options

Accessories

All technical data is indicated in the instructions for use.

The functions of the individual options are detailed in the instructions for use.

Perfusor® Space

Designation Part No.:

Power supply Euro . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0871 3110A Power supply UK. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0871 3111A Power supply USA / Japan . . . . . . . . . . . . . . . . . . . . . . 0871 3112A Power supply Australia. . . . . . . . . . . . . . . . . . . . . . . . . 0871 3113A Power supply South Africa. . . . . . . . . . . . . . . . . . . . . . 0871 3115A

Designation

Part No.:

Charger SP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

0871 3170

battery charging station

Connection cable staff call SP. . . . . . . . . . . . . . . . . . . .

0871 3232

Power supply cable 12 V . . . . . . . . . . . . . . . . . . . . . . . .

0871 3231

for ambulance cars

CombiLead SP 12 V . . . . . . . . . . . . . . . . . . . . . . . . . . . .

0871 3133

connection cable, pump — pump

InterfaceLead SP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

0871 3234

interface cable RS232

InterfaceLead SP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

0871 3230

interface cable CAN SP

SpaceClamp SP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

0871 3130

The SpaceClamp is a holder attached on beds for one

or several Space system pumps.

Short stand SP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

0871 3135

Space PCA kit (PCA button) . . . . . . . . . . . . . . . . . . . . .

0871 3554

Syringe Anti Removal Cap PSP . . . . . . . . . . . . . . . . . . .

0871 3556

1 — 12

Perfusor® Space, 1.3 gb

Unit Diagnosis / Calibration 2

General

WARNING

WHILE TESTING THE UNIT AND TROUBLE SHOOTING THE

OPERATOR/SERVICE TECHNICIAN MUST WORK WITH VOLTAGES

UP TO 115 / 230 V AC. THESE VOLTAGES MAY CAUSE INJURIES

WHICH ARE DANGEROUS TO LIFE AND LIMB. THE NATIONAL AND

INTERNATIONAL SAFETY REGULATIONS ARE TO BE ADHERED TO.

Before each disassembly and assembly of a unit subsystem check the connectors, plug contacts and connections for corrosion and tight fit. These fault types are not described again in the following trouble shooting list.

The following equipment and gauges are necessary for testing the unit and/or performing troubleshooting:

— PC

— Service connector SP

— Service Program HiBaSeD — Interface cable

— Syringe 2 ml / 3 ml — Syringe 10 ml

— Syringe 30 ml

— Diameter gauge 32.0 mm — Diameter gauge 23.4 mm — Diameter gauge 15.7 mm — Diameter gauge 9.0 mm — Length gauge PSP

— Syringe gauge “#Lehre OPS 50“with push-button plate and motor power test adapter for Perfusor® Space

There are pictures of the gauges in Chapter “Special Tools“ ( pg. 8 — 3).

CAUTION

Take special care when carrying out measurements on an open and switched-on unit. Short circuits and wrong measuring methods can cause serious damage to or destroy the subsystems of the device.

Perfusor® Space, 1.1 gb

2 — 1

2

Unit Diagnosis / Calibration

The unit check, calibration and trouble shooting are subdivided into numbered working steps (Unit Test Step UTS, Calibration Step CS, Trouble Shooting TS) and are based on each other.

Beginning with UTS 1 the operation described here has to be executed. The consequences of the steps performed are listed in the “Function“ column. If the result corresponds to the consequence, the working step must be carried out to which reference is made in the column “If yes”. If the result does not correspond with the function described, the working step in column “If no” is to be executed.

One example is given in Fig.: 2 — 1.

UTS

Activity

Function

If yes

If no

1

UTS 2

2

UTS 3

TS 1

3

UTS 4

4

UTS 5

TS 4

5

Model table 1

TS

Activity

Function

1

UTS 3

TS 2

2

TS 3

TS 4

3

UTS 3

4

UTS 4

TS 4

5

UTS 4

Model table 2

Fig.: 2 — 1 Model tables

Steps for which additional information is required are described after the table in detail.

2 — 2

Perfusor® Space, 1.0 gb

Unit Diagnosis / Calibration

2

Alarms and Error Codes

The alarms of the Perfusor® Space are classified in 5 categories. These categories are listed hereafter according to their importance.

Alarm advice

In case of unacceptable inputs corresponding messages are displayed (e.g. “Caution! Rate out of range“, “The parameter cannot be changed“) and a beep sounds.

Pre-alarm

Pre-alarms are triggered several minutes (depending on the service settings) before the operating alarms.

Reminder alarm

A reminder alarm is triggered if the device is not operated for two minutes when input or operation was not finished.

Operating alarm

In case of an operating alarm the infusion is stopped. An audible signal is released, the red LED flashes and a staff call is triggered. The message “Alarm” and the cause of the alarm appear on the display.

Device alarm

The most important alarms and error codes as well as their meaning and possible fault clearance are specified in the following lists.

Note

The device should be checked after every repair or service (see “Device Check“ pg. 2 — 9).

Perfusor® Space, 1.0 gb

2 — 3

2

Unit Diagnosis / Calibration

Alarms

Alarm

Possible Cause

Fault Clearance

1

Battery nearly discharged (type: pre-

The device was not connected to the

Operate the device with battery until the

alarm)

mains long enough

message “Battery discharged“ is

displayed and the unit is switched off.

Then connect the unit to the mains for at

least 6 hours.

Battery module defective or too old

Replace battery module

2

Battery discharged (type: operating

The device was not connected to the

Connect the unit to the mains for at least

alarm)

mains long enough

6 hours

Battery module defective

Replace battery module

3

Battery cover open (type: operating

The battery compartment cover is

Insert the battery compartment cover

alarm)

not correctly closed

correctly

The magnet in the battery

Exchange the battery compartment cover

compartment cover is missing

The battery compartment cover is

Replace battery module

not recognized by the battery

module

4

Drive blocked (type: operating alarm)

The drive was manually blocked

Eliminate blockage

Driving force too low

Connect the unit to the mains for at least

6 hours and charge battery

Re-calibrate the device

The drive is physically damaged

Replace drive.

5

Malfunction of claws (type: operating

The syringe piston was not

Select or insert correct syringe type

alarm)

recognized

Loosen the syringe via the emergency

release button in the drive head and

insert again

Re-calibrate the device

The claws or the claw drive are/is

Replace drive head

damaged

6

Push-button has no contact (type:

Negative pressure in the syringe

See instructions for use

operating alarm)

system

Syringe was removed without

See instructions for use

opening the syringe holder

Push-button sensor defective

Replace drive head

7

Device alarm (type: device alarm)

A serious internal fault was detected in

Switch device off and on

the system

Carry out a device check (see “Device

Check“ pg. 2 — 9)

Table 2 — 1 Alarms

2 — 4

Perfusor® Space, 1.0 gb

Unit Diagnosis / Calibration

2

Device Alarms of the Function Processor

Error Code

Definition

Possible Cause

Fault Clearance

1

1001 … 1013

Internal Error

2

1014

Loudspeaker not off

Loudspeaker connector

Check the loudspeaker connector

Loudspeaker

Check the loudspeaker

3

1015

Loudspeaker lost

Loudspeaker connector

Check the loudspeaker connector

Loudspeaker

Check the loudspeaker

4

1016

Loudspeaker shorted

Loudspeaker connector

Check the loudspeaker connector

Loudspeaker

Check the loudspeaker

5

1017

KuP switchoff path defect

Switch off path

(K_SM_CLK)

6

1018

ADC pressure out of range

Pressure measurement in drive

Carry out calibration

head

7

1019

Internal Error

8

1020

FUP Flash Memory Error Software

Software

Update unit software

9

1021

FUP different version KuP to FuP

Software

Update unit software

10

1022

FUP pressure zero test fail

Pressure measurement in drive

Carry out calibration

head

11

1023

FUP pressure offset test fail

Pressure measurement in drive

Carry out calibration

head

12

1024

FUP EA key closed too long 20sec

Keyboard defective

Carry out device check

13

1025

Internal Error

Table 2 — 2 Device alarms of the function processor

Perfusor® Space, 1.0 gb

2 — 5

2

Unit Diagnosis / Calibration

Device Alarms of the Control Microprocessor

Error Code

Definition

Possible Cause

Fault Clearance

1

1100

Timebase too fast

Quartz of the processor PCB

Exchange processor PCB

2

1101

Timebase too slow

Quartz of the processor PCB

Exchange processor PCB

3

1102

Timebase fail

Quartz of the processor PCB

Exchange processor PCB

4

1103

Keyboard High

Keyboard defective

Carry out device check

5

1104

EA_KEY defect 25sec

Keyboard defective

Carry out device check

6

1105

No keydecode

Keyboard defective

Carry out device check

7

1106

ROM Romtest defect Software

Software

Update unit software

8

1107

ROM Program defect

Software

Update unit software

9

1108

CM State without set

K_V_KM_ON

10

1109

MPU_Test failed

Software

Update unit software

11

1110

RAM_Test failed

Software

Update unit software

12

1111

active reset

Voltage supply during operation

interrupted

13

1112 … 1114

Internal Error

14

1115

Drive too fast

Motor drive

Exchange processor PCB

Recognition of direction of

rotation

15

1116

Drive too slow

Motor drive

Exchange processor PCB

Recognition of direction of

rotation

16

1117 … 1118

Internal Error

17

1119

lcd backlight on defect

LC display defective

Exchange operating unit

18

1120

lcd backlight off defect

LC display defective

Exchange operating unit

19

1121

red led on defect

LC display defective

Exchange operating unit

20

1122

red led off defect

LC display defective

Exchange operating unit

21

1123

key pressed too long (without EA-

Keyboard defective

Carry out device check

Key) 60sec

22

1124 … 1127

Internal Error

23

1128

Drive motion rightless forward

Motor drive

Exchange processor PCB

Recognition of direction of

rotation

24

1129

Drive motion rightless backward

Motor drive

Exchange processor PCB

Recognition of direction of

rotation

25

1130 … 1200

Internal Error

Table 2 — 3 Device alarms of the control microprocessor (Part 1 of 2)

2 — 6

Perfusor® Space, 1.0 gb

Unit Diagnosis / Calibration

2

Error Code

Definition

Possible Cause

Fault Clearance

26

1201

different version FuP to KuP

Software

Update unit software

Software

27

1202

E_ERROR_STEPMOTOR_1 Phase

Drive motor, lead screw

Exchange processor PCB

not ok

28

1203

E_ERROR_STEPMOTOR_2 Current

Motor drive

Carry out calibration

value not 0x55

Recognition of direction of

rotation

29

1204

E_ERROR_STEPMOTOR_3

Motor drive

Carry out calibration

K_SM_CLK defect

Recognition of direction of

rotation

30

1205

E_ERROR_STEPMOTOR_4 Phase

Motor drive

Carry out calibration

not ok

Recognition of direction of

rotation

31

1206

E_ERROR_STEPMOTOR_5

Motor drive

Carry out calibration

Current value not 0

Recognition of direction of

rotation

32

1207

E_ERROR_STEPMOTOR_6 Current

Motor drive

Carry out calibration

value not 0x55

Recognition of direction of

rotation

33

1208

E_ERROR_STEPMOTOR_7 Current

Motor drive

Carry out calibration

value not 0xAA

Recognition of direction of

rotation

34

1209

E_ERROR_STEPMOTOR_8 Phases

Motor drive

Carry out calibration

not 0

Recognition of direction of

rotation

35

1210

E_ERROR_DCMOTOR_1

Piston brake drive motor def.

Claw drive in drive head defective

36

1211

E_ERROR_DCMOTOR_2

Piston brake light barrier def.

37

1212

E_ERROR_DCMOTOR_3

38

1213

E_ERROR_DCMOTOR_4

39

1214

E_ERROR_DCMOTOR_5

40

1215

no V_MOT

Voltage transformer defective

Exchange processor PCB

41

1216

overvoltage test fail

42

1217

no V_MOT

43

1218

undervoltage test fail

44

1220

syringeholder defect

Syringe holder or potentiometer

Replace syringe holder

def.

Exchange processor PCB

45

1221

syringe change timeout

46

1237 … 1238

Internal Error

47

1239

plunger plate sensor defect

Pressure measurement in drive

Replace drive head

head

48

1240 … 1254

Internal Error

Table 2 — 3 Device alarms of the control microprocessor (Part 2 of 2)

Perfusor® Space, 1.0 gb

2 — 7

2

Unit Diagnosis / Calibration

The Most Important Error Modes

The following list specifies the most important error modes and

their clearance.

Note

The device must be checked after every repair or service (see

“Device Check“ pg. 2 — 9).

Error

Possible Cause

Fault Clearance

1

The battery module discharges too fast

The device was not used for a longer time.

Discharge and charge battery module

The battery module was not discharged

several times

and charged at regular intervals.

Replace battery module

Table 2 — 4

2 — 8

Perfusor® Space, 1.0 gb

Unit Diagnosis / Calibration

2

Device Check

UTS

Activity

Function

If yes

If no

1

The device is inserted in a SpaceStation or

UTS 2

UTS 3

connected to a SpaceControl.

2

Remove the device.

UTS 4

3

Loosen all connections from the device.

UTS 4

4

Remove syringe and close syringe holder.

UTS 5

5

Plug service connector SP on connector “P2”.

UTS 6

6

Connect power supply to the device via service

All LEDs light up for a short moment.

UTS 7

TS 1

connector SP.

7

The battery charge state and the mains connection

UTS 8

TS 5

are displayed at the top left of the LC display

(without lighting).

8

Switch on unit.

All LEDs light up (from left: yellow, green, blue).

UTS 9

TS 5

9

A short deep and then a short high beep sound.

UTS 10

TS 7

10

The colour of the middle LED changes from green to

UTS 11

TS 8

red, then the LED goes out. The yellow and the blue

LED remain on for a short moment.

11

The message “Self-test active” and the current

UTS 12

TS 8

software version are displayed.

12

Keyboard, LC display as well as the syringe area are

UTS 13

TS 9

illuminated.

13

The drive head moves to the extended end position.

UTS 14

TS 11

14

The claws in the syringe head close and open.

UTS 15

TS 14

15

The message “Drive moves back / Syringe change”

UTS 16

TS 16

appears on the display.

16

“Open syringe holder and insert syringe or press “C“

UTS 17

TS 16

to input parameters“ is displayed.

17

Open syringe holder.

“Syringe change / Please insert syringe …” is

UTS 18

TS 17

displayed.

18

Press the “>” key.

The service information:

UTS 19

TS 19

— Brake: not started or active

— drivetest ok

— Size: 35.4 KuP 35.4 FuP

is displayed on the LC display.

19

Insert syringe 30 ml.

The syringe piston is fastened with the syringe

UTS 20

TS 21

holder blade.

20

On the LC display “Brake: stopped by current”

UTS 21

TS 21

appears in the line.

Table 2 — 5 Device check

(Part 1 of 5)

Perfusor® Space, 1.1 gb

2 — 9

2

Unit Diagnosis / Calibration

UTS

Activity

Function

If yes

If no

21

Insert 2 ml / 3 ml syringe.

On the LC display “Brake: stopped by holder”

UTS 22

TS 21

appears in the line.

22

Open syringe holder and remove syringe.

On the LC display “Brake: stopped by light barrier”

UTS 23

TS 21

appears in the line.

23

3.54 is shown for FuP on the LC display. The value

UTS 24

CS 1

displayed for FuP may have a maximum tolerance of

± 0.04.

24

Close syringe holder.

The value for FuP changes to 7.0 ± 0.4.

UTS 25

CS 1

25

Insert diameter gauge 9.0 mm.

The value for FuP changes to 9.0 ± 0.4.

UTS 26

CS 1

26

Insert diameter gauge 15.7mm.

The value for FuP changes to 15.7 ± 0.4.

UTS 27

CS 1

27

Insert diameter gauge 23.4 mm.

The value for FuP changes to 23.4 ± 0.4.

UTS 28

CS 1

28

Insert diameter gauge 32.0 mm.

The value for FuP changes to 32.0 ± 0.4.

UTS 29

CS 1

29

The sum of the tolerances of UTS 23 to UTS 28 must

UTS 30

CS 1

not exceed 1.0.

30

Insert 2 ml / 3 ml syringe.

UTS 31

31

Press the “>” key.

The syringe selection is displayed.

UTS 32

32

Select a syringe.

The drive head moves to the syringe piston, the

UTS 33

TS 26

claws in the drive head close and the message

“Syringe is caught / Please wait” is displayed.

33

Test all buttons on the operating unit during a

When the buttons are pressed the desired reaction

UTS 34

TS 29

functional check (carry out infusion).

is carried out.

34

Open syringe holder while the infusion is

The red LED on the operating unit flashes and the

UTS 35

TS 31

administered.

red LED of the service connector SP lights up. The

message “Alarm / Syringe holder” is displayed.

35

Close syringe holder and continue infusion.

UTS 36

36

Stop infusion.

UTS 37

37

Open syringe holder.

“Syringe change / Initiate change? Yes / No” is

UTS 38

displayed.

38

Confirm with “Yes”.

The claws in the drive head open and the drive head

UTS 39

moves to the extended end position.

39

Remove syringe.

UTS 40

Table 2 — 5 Device check (Part 2 of 5)

2 — 10

Perfusor® Space, 1.1 gb

Unit Diagnosis / Calibration

2

UTS

Activity

Function

If yes

If no

40

Insert syringe gauge for the strain gauge

UTS 41

measurement, close syringe holder and select

syringe type „#Lehre OPS50“. The syringe gauge

must not be tipped. Therefore fix the syringe gauge

so far into the syringe recess by hand that the

piston brake moves back and the claws surrounds

the pressure element.

WARNING

DURING THE STRAIN GAUGE MEASUREMENT WITH

SYRINGE GAUGE THE SYRINGE HOLDER MUST NOT

BE OPENED. THE SYRINGE GAUGE IS UNDER VERY

HIGH PRESSURE AND MAY CAUSE INJURIES IF THE

PRESSURE IS RELIEVED SUDDENLY.

41

Input a delivery rate of 200 ml/h, select pressure

When the maximum pressure of this pressure stage

UTS 42

CS 1

stage 1 and start infusion.

is reached, the delivery is stopped, the red LED on

the operating unit flashes and the message “Alarm

/ Pressure too high” is displayed.

The value read on the syringe gauge (in N) must

match the value indicated for the strain gauge

measurement of this pressure stage in the TSC.

42

Confirm alarm.

UTS 43

43

Select pressure stage 3 and start infusion.

When the maximum pressure of this pressure stage

UTS 44

CS 1

is reached, the delivery is stopped, the red LED on

the operating unit flashes and the message “Alarm

/ Pressure too high” is displayed.

The value read on the syringe gauge (in N) must

match the value indicated for the strain gauge

measurement of this pressure stage in the TSC.

44

Confirm alarm.

UTS 45

45

Select pressure stage 8 and start infusion.

When the maximum pressure of this pressure stage

UTS 46

CS 1

is reached, the delivery is stopped, the red LED on

the operating unit flashes and the message “Alarm

/ Pressure too high” is displayed.

The value read on the syringe gauge (in N) must

match the value indicated for the strain gauge

measurement of this pressure stage in the TSC.

46

Confirm alarm and pull syringe holder briefly.

UTS 47

Table 2 — 5 Device check (Part 3 of 5)

Perfusor® Space, 1.3 gb

2 — 11

2

Unit Diagnosis / Calibration

UTS

Activity

Function

If yes

If no

47

Confirm syringe change, release syringe gauge and

UTS 48

remove gauge.

WARNING

WHILE CHECKING THE MOTOR POWER LIMITATION

WITH THE SYRINGE GAUGE THE SYRINGE HOLDER

MUST NOT BE OPENED. THE SYRINGE GAUGE IS

UNDER VERY HIGH PRESSURE AND MAY CAUSE

INJURIES IF THE PRESSURE IS RELIEVED SUDDENLY.

48

Insert the motor power test adapter in the drive

UTS 49

head to check the motor power limitation.

49

Dismount the push-button plate from the syringe

UTS 50

gauge and insert syringe gauge.

50

Select syringe type “#Lehre OPS 50“. The threaded

UTS 51

end of the syringe gauge must be introduced in the

opening of the motor power test adapter. To do this,

hold on to the syringe gauge, if necessary by hand,

in the syringe area.

51

Select pressure stage 1 and start infusion.

When the maximum pressure of this pressure stage

UTS 52

CS 1

is reached, the delivery is stopped, the red LED on

the operating unit flashes and the message “Alarm

/ Drive blocked” is displayed.

The value read on the syringe gauge (in N) must

match the value indicated for the motor power

limitation in the TSC.

52

Confirm alarm.

UTS 53

53

Select pressure stage 3 and start infusion.

When the maximum pressure of this pressure stage

UTS 54

CS 1

is reached, the delivery is stopped, the red LED on

the operating unit flashes and the message “Alarm

/ Drive blocked” is displayed.

The value read on the syringe gauge (in N) must

match the value indicated for the motor power

limitation in the TSC.

54

Confirm alarm.

UTS 55

Table 2 — 5 Device check (Part 4 of 5)

2 — 12

Perfusor® Space, 1.0 gb

Unit Diagnosis / Calibration

2

UTS

Activity

Function

If yes

If no

55

Select pressure stage 6 and start infusion.

When the maximum pressure of this pressure stage

UTS 56

CS 1

is reached, the delivery is stopped, the red LED on

the operating unit flashes and the message “Alarm

/ Drive blocked” is displayed.

The value read on the syringe gauge (in N) must

match the value indicated for the motor power

limitation in the TSC.

56

Confirm alarm and pull syringe holder briefly.

UTS 57

57

Confirm syringe change, release syringe gauge and

UTS 58

remove gauge.

WARNING

WHILE CHECKING THE MOTOR POWER LIMITATION

WITH THE SYRINGE GAUGE THE SYRINGE HOLDER

MUST NOT BE OPENED. THE SYRINGE GAUGE IS

UNDER VERY HIGH PRESSURE AND MAY CAUSE

INJURIES IF THE PRESSURE IS RELIEVED SUDDENLY.

58

Insert syringe type 50/60 ml and lock PCA-lock with

The syringe holder cannot be opened.

UTS 59

TS 32

PCA-key.

59

Open PCA-lock and remove syringe.

UTS 60

60

Insert syringe type 10 ml and lock PCA-lock with

The syringe holder cannot be opened.

UTS 61

TS 32

PCA-key.

61

Open PCA-lock and remove syringe.

UTS 62

62

Switch device off.

The message “Pump is switched off in 3 .. 2 .. 1 sec”

UTS 63

is displayed.

63

“Drive is parked …/ Please wait …“ is displayed. The

UTS 64

drive head moves to the retracted park position.

64

The device switches off.

UTS 65

TS 35

65

Pull off the power supply.

The blue LED lights up for a short moment.

UTS 66

TS 35

66

Switch on unit.

UTS 67

67

Open the battery compartment cover when the

An alarm signal sounds, the red LED flashes and

UTS 68

TS 36

drive head has moved to the extended end position.

“Alarm / Battery cover open / Confirm with “OK” is

displayed.

68

Remove battery.

A permanent alarm is triggered.

UTS 69

TS 38

69

Insert battery, close battery compartment cover and

The message “Devicealarm / 1111” is displayed.

UTS 70

switch on the device.

70

Switch the device off, remove service connector SP

This step

terminates

and dismount test structure.

the device check.

Table 2 — 5 Device check (Part 5 of 5)

Perfusor® Space, 1.3 gb

2 — 13

2

Unit Diagnosis / Calibration

Calibration

CS

Activity

Function

If yes

If no

1

Connect unit to PC with interface cable.

CS 2

2

Start Service Program on the PC (see “Starting the

The desired device is found by the Service Program

CS 3

Service Program“ pg. 1 — 7).

and then displayed.

3

Start calibrating the unit (see “Starting Calibration“

CS 4

pg. 2 — 14).

4

Carry out calibration of the claws (see “Claw

Calibration of the claws was terminated

CS 5

Calibration“ pg. 2 — 23).

successfully.

5

Carry out calibration of the syringe holder (see

Calibration of the syringe holder was terminated

CS 6

“Syringe Holder Calibration“ pg. 2 — 23).

successfully.

6

Carry out pressure calibration (see “Pressure

Pressure calibration was terminated successfully.

CS 7

Calibration“ pg. 2 — 26).

7

Close the Service Program (see “Quit the Service

UTS 23

Program“ pg. 1 — 11).

Table 2 — 6 Calibration

Procedural Instructions for Calibration

Starting Calibration

Note

Calibration must be carried out with power supply connected, since the calibration can be interrupted suddenly if the unit is battery-operated and the battery gets discharged so that the device is switched off.

Note

Please note that text and / or functions of the Service Program may change depending on the software version. The following screen illustrations are only examples and represent the state when the manual was printed.

2 — 14

Perfusor® Space, 1.1 gb

Unit Diagnosis / Calibration

2

1. Start the Service Program (see “Starting the Service Program“pg. 1 — 7).

2. Select the unit to be calibrated in the left column of the window with a double mouse-click. The blue and the yellow LED blinks in opposite with the red LED.

3. Select the register tab “Calibration”.

Fig.: 2 — 2

4. Press the “New device” button in the frame “Calibration procedure”. The window “Worker ID” is opened.

Note

If you do not have an allocated worker id, enter “0001”.

Perfusor® Space, 1.2 gb

2 — 15

2

Unit Diagnosis / Calibration

Fig.: 2 — 3

Fig.: 2 — 4

Fig.: 2 — 5

5.Input your user number in the window “Worker ID” as well as the six-digit serial number of the device, if necessary.

6.Confirm the input with “OK”.

Note

If HiBaSeD could not clearly read the device serial number, the number must be entered according to the rating plate.

If the unit is not yet switched on, the window “Device on/off” opens and the user is asked to select the desired language.

7.Select the desired language. The respective operating steps are explained in detail in the instructions for use.

After the language was confirmed the unit switches on and the window “Device on/off” closes.

Note

If calibration is interrupted, data is written back to the device and marked as invalid if this is still possible. When the Service Program is started again the data is marked as faulty and highlighted red when the “Modify” button is selected in the “Calibration procedure” frame.

2 — 16

Perfusor® Space, 1.2 gb

Loading…

You can only view or download manuals with

Sign Up and get 5 for free

Upload your files to the site. You get 1 for each file you add

Get 1 for every time someone downloads your manual

Buy as many  as you need

Braun Perfusor Space Instructions For Use Manual

  1. Manuals
  2. Brands
  3. Braun Manuals
  4. Health Care Products
  5. Perfusor Space
  6. Instructions for use manual

  • Contents

  • Table of Contents

  • Bookmarks

Quick Links

Perfusor® Space

and Accessories

Instructions for Use

It is recommended that all pumps at

your care unit are equipped with the

same software version.

GB

Valid for software 688M

loading

Related Manuals for Braun Perfusor Space

  • Medical Equipment Braun Perfusor Space Instructions For Use Manual

Summary of Contents for Braun Perfusor Space

  • Page 1
    Perfusor® Space and Accessories Instructions for Use It is recommended that all pumps at your care unit are equipped with the same software version. Valid for software 688M…
  • Page 2: Table Of Contents

    CONTENTS Perfusor® Space Overview………………………3 Symbols on Product ……………………….5 Patient Safety ……………………….6 Menu Structure / Navigation……………………11 Chapter 1 Operation ……………………..14 1.1 Start of Infusion …………………….14 1.2 Entry With Different Combinations of Rate, VTBI (= Volume To Be Infused) and Time……………………….15 1.3 Bolus Application ……………………16 1.4 Syringe Change and New Therapy Start …………….17 1.5 End of Infusion……………………..18 1.6 Standby Mode ……………………..18…

  • Page 3: Perfusor® Space Overview

    PERFUSOR SPACE® OVERVIEW P E R F U S O R ® S PA C E O V E R V I E W Arrow up and -down Press to reset single values Drive head with Scroll through menus, change setting of numbers from 0-9,…

  • Page 4
    For vertical fixation of PoleClamp push lever down and rotate either way until lever clicks into notch. Push lever for rotation. Caution: A maximum of three B. Braun Space pumps can be stacked together when used with the PoleClamp SP.
  • Page 5: Symbols On Product

    SYMBOLS ON PRODUCT S Y M B O L S O N P R O D U C T Symbol Explanation Mandatory action: see instruction for use. See accompanying documents. Type CF unit with defibrillation protection Protection class II device Symbol indicating separate collection for electrical and electronic equipment (2002/96/EC) CE mark compliant to Directive 93/42/EEC…

  • Page 6: Patient Safety

    Operation • The initial training of the Perfusor® Space is to be performed by B. Braun sales personnel or other authorized persons. After each software update, the user is required to inform himself of the changes to the device and accessories by referring to the Instructions for Use.

  • Page 7
    PATIENT SAFETY • Only connect to patient once the syringe has been inserted correctly and there is proper fixation of the syringe pressure plate by the claws of the drive head. Interrupt connection during syringe change to prevent incorrect dose delivery.
  • Page 8
    PATIENT SAFETY Enteral Nutrition The Perfusor® Space may be used for enteral nutrition. Do not use enteral fluids for intravenous infusion as this may harm your patient. For this reason only use disposables dedicated and labeled for enteral nutrition. Other components •…
  • Page 9
    TCI and properly trained in using the present device. • The use of TCI with B. Braun Space does not limit the responsibility of the anaesthetist for administration of drugs. They need to be fully aware of the available literature for any parameter set used in association with a drug and need to refer to the prescribed information for rate and dosing limits.
  • Page 10
    PATIENT SAFETY • While using TCI an appropriate patient monitoring is mandatory. • Take care of using the right dilution/concentration of the drug and make sure the right dilution is selected at the pump. • Never administer Propofol or Remifentanil by a second infusion as long as you use TCI.
  • Page 11: Menu Structure / Navigation

    MENU STRUCTURE / NAVIGATION MENU STRUCTURE / NAVIGATION Cutline On/Off button OK button Start/Stop button Keypad with arrow up, -down, -left, -right button Bolus button Connection button Clear button All display screen shots are examples and may be different when related to an individual patient and individualized therapy.

  • Page 12
    MENU STRUCTURE / NAVIGATION Display Meaning All status information is available in the bottom line of the dislplay. The desired information can be selected by using and will be displayed permanently thereafter (e. g. drug long name, time until syringe empty, current system pressure etc.).
  • Page 13
    MENU STRUCTURE / NAVIGATION Start Up Main Special Options Status Menu Menu Functions Menu Menu Syringe Dose Rate Occlusion Intermediate Dose selection Calculation Pressure volume Intermediate Prime ? Concentration Drug Library OccluGuard amount Change-over Use last Pressure Intermediate Weight therapy ? from Leap/Drop time…
  • Page 14: Chapter 1 Operation

    OPERATION Chapter 1 OPERATION 1.1 Start of Infusion • Ensure correct installation of the pump device. If the pump is connected to mains, the display states information such as the battery status, the mains connection symbol and the last therapy. •…

  • Page 15: Entry With Different Combinations Of Rate, Vtbi (= Volume To Be Infused) And Time

    OPERATION Chapter 1 • Press to commence infusion. Running arrows on display and green LED above display indicate pump is infusing. Note: Stop the infusion at any time by pressing . The pump can be turned off at any time by pressing for 3 sec (Exception: Data lock level 2) and as long a disposable is inserted.

  • Page 16: Bolus Application

    OPERATION Chapter 1 a) Target symbol is placed in front of VTBI: • Change of VTBI => Adjustment of time. Old and new target: VTBI • Change of time => Adjustment of rate. Old and new target: VTBI b) Target symbol is placed in front of time: •…

  • Page 17: Syringe Change And New Therapy Start

    OPERATION Chapter 1 1.4 Syringe Change and New Therapy Start Note: To avoid incorrect dosing, always disconnect the pump from the patient when changing the syringe. Never leave the pump device unattended during syringe change. Before inserting a new syringe check if the axial fixation is properly working. •…

  • Page 18: Standby Mode

    OPERATION Chapter 1 • Open pump cover. Remove the syringe, move the syringe holder into an upright position and close the front door. • Press for 3 sec. to switch the pump off. The drive moves into parking position. Note: The settings will be permanently saved by the switched off device.

  • Page 19: Chapter 2 Advanced Operations

    OPERATION Chapter 1 ADVANCED OPERATIONS 2.1 Status Request of Pump when Infusion is Running Press to switch between run display and Main Menu while the device is infusing. Navigate through the menu using to check parameters. In order to check the menu parameters in the Status-/Options Menu, select «Status»…

  • Page 20: Chapter 3 Special Functions

    ADVANCED OPERATIONS Chapter 2 SPECIAL FUNCTIONS 3.1 Dosing Units and Dose Rate Calculation (Overview) The following list shows the units used in the pump: Gram family: ng, mcg, mg, g Unit family: mIU, IU, kIU, MIU Equivalents family: mEq Mole family: mmol Kilocalorie family: kcal Millitliter family:…

  • Page 21: Dose Rate Calculation (Operation)

    SPECIAL FUNCTIONS Chapter 3 3.2 Dose Rate Calculation (Operation) a Select dose rate calculation with l. a Select the unit of the active ingredient with and confirm it with l. a Enter the concentration by entering the amount of the active ingredient and the volume.

  • Page 22
    SPECIAL FUNCTIONS Chapter 3 On the one hand, a drug name including the according therapy data can be taken from the drug library. On the other hand, if a rate, VTBI and/or time were already defined in the Main Menu, the drug name and the adjusted values of the data set will be loaded.
  • Page 23
    SPECIAL FUNCTIONS Chapter 3 • Select the desired drug with and press l. Before the initial bolus begins, the bolus menu is displayed to allow editing the bolus with q. • Check the parameter and start infusion with Hard Limits: If the set rate/dose/bolus volume and bolus rate exceed the values stored in the drug library (hard limits), the drug will be rejected, a hint will be displayed and the pump will fall back into the drug selection.
  • Page 24: Patient Controlled Analgesia (Pca) (Optional)

    SPECIAL FUNCTIONS Chapter 3 The Drug Library Upload starts as soon as the pump is in Passive mode. You can cancel the upload by pressing c. Note: Please contact your local sales represantative in case you like to use Remote Drug Library update.

  • Page 25
    SPECIAL FUNCTIONS Chapter 3 In this state the patient is allowed to demand boli. Depending on the status of the therapy these are either administered or denied. Changing the syringe is also possible by using the code for level 1 or level 2. Altering the settings for PCA or other therapies however is only possible with the code for level 3.
  • Page 26: Target Controlled Infusion (Tci) (Optional)

    The pharmacokinetic model and its parameters are schematically depicted by the following illustration: B. Braun Space is offering two modes for TCI: • TCI by targeting the plasma concentration In this mode the user selects the desired concentration of a drug in the…

  • Page 27
    TCI. A pharmacokinetic model modi- fied in such way is schematically depicted by the illustration on the next page. TCI with B. Braun Space is possible with two drugs: Propofol and Remifentanil. For Propofol the user can choose between two parameter sets. The parameter…
  • Page 28
    SPECIAL FUNCTIONS Chapter 3 Drug / Parameter Propofol Remifentanil [Litre] 0,228 * Weight 4,27 5,1 — 0,0201 * (Age — 40) + 0,072 * (LBM — 55) [min 0,119 0,443 + 0,0107 * (Weight — [2,6 — 0,0162 * (Age — 40) + 77) — 0,0159 * (LBM — 59) + 0,0191 * (LBM — 55)] / [5.1 — 0,0062 * (Height — 177)
  • Page 29
    Chapter 3 Important note: Before installing an additional drug list please contact your local B. Braun representative! Setting up the pump For TCI a drug list with at least one drug activating the profile TCI is necessary. The drug list in this version is pre-defined. By this the conditions for an effective and safe therapy are defined.
  • Page 30
    SPECIAL FUNCTIONS Chapter 3 Important notes: • Be sure to enter the data corresponding to the respective patient. • Once the TCI is started patient data can not be altered! Editing a target and starting TCI The editor window for setting the target comes up with the default value from the drug list.
  • Page 31
    SPECIAL FUNCTIONS Chapter 3 Useful information while pump is running By pressing additional information can be requested. Pressing a second time is offering a graphical overview. The line describes the course of Cp over the time and the area describes the course of Ce over the time.
  • Page 32: Barcoding

    SPECIAL FUNCTIONS Chapter 3 Barcoding The barcoding functionality is included but initially not activated. Please contact your local sales representative in case you like to use barcoding. Ramp and Taper Mode The Ramp and Taper Mode is designed to deliver infusions with gradual ramp up and taper down rates.

  • Page 33
    SPECIAL FUNCTIONS Chapter 3 Set Profile Parameters: The therapy can be started directly via the drug library or via the Main Menu/Special functions. Starting Ramp and Taper via Drug Library: Note: Ramp and Taper settings have been configured in the Drug List Manager before and have been uploaded into the pump.
  • Page 34: Program Mode

    SPECIAL FUNCTIONS Chapter 3 Taper phase The pump linearly decreases the rate in the predefined time until it reaches the KVO rate Note: After starting infusion it is only possible to change rates, time and VTBI in the continuous phase. By editing (increasing/decreasing) the plateau rate, the therapy is recalculated.

  • Page 35
    SPECIAL FUNCTIONS Chapter 3 Example: Program Mode should only be performed by an experienced user being familiar with the principles of the Program Mode function and properly trained in using the present device. Note: The active Program Mode function always displays this icon in the Display Note: Bolus function is disabled for Program Mode.
  • Page 36: Intermittent Mode

    SPECIAL FUNCTIONS Chapter 3 The pump can be started now by pressing Starting Program Mode via Special Function Menu: • Switch on pump with and wait until self-check is finished. • Insert disposable. • Go to Special Functions Menu and select Program Mode. •…

  • Page 37
    SPECIAL FUNCTIONS Chapter 3 Example: Intermittent Mode should only be performed by an experienced user being familiar with the principles of the Intermittent Mode and properly trained in using the present device. Note: The active Multi Dose Mode function always displays this icon in the Display Note: Regular Bolus function is disabled for Intermittent Mode.
  • Page 38
    SPECIAL FUNCTIONS Chapter 3 Starting Intermittent Mode via Special Function Menu: • Switch on pump with and wait until self-check is finished. • Insert disposable. • Go to Special Functions Menu and select Intermittent Mode. • Press to enter parameters and to confirm.
  • Page 39: Dose Over Time

    Note: The feature Dose Over Time always requires the usage of dosing units (i.e., mg or mg/kg patient weight). Before using Dose Over Time contact your local B. Braun representative! Starting Dose Over Time via Drug Library: Note: Dose Over Time settings have been configured in the Drug List Manager before and have been uploaded into the pump.

  • Page 40
    SPECIAL FUNCTIONS Chapter 3 • Insert disposable and use the drug library according to the Instructions for Use. • Select a drug by using and press l. The pump now offers the possible therapy profiles. Select “Dose over Time” with press l.
  • Page 41: Take Over Mode (Tom) (Optional)

    SPECIAL FUNCTIONS Chapter 3 3.11 Take Over Mode (TOM) Take Over Mode is a feature to support the user during syringe changes by auto- matically starting a second Perfusor® Space pump when the first has run empty. The second pump automatically takes over the infusion rate from the first pump. Activation: •…

  • Page 42
    SPECIAL FUNCTIONS Chapter 3 • Navigate through the list with and select in alphabetical order (all drugs) or within a category with l. The drug selected in the second pump must be the same as the first. • Navigate through the list with and select a concentration with l.
  • Page 43
    SPECIAL FUNCTIONS Chapter 3 Note: Start-up behaviour is not influenced by TOM. See Chapter Start Up Graphs and Trumpet Curves. Note: Please use a seperate patient connection for Take Over Mode infusion (e.g. smallbore extension set) or use a back check valve for lines at the same access which are not used for Take Over Mode.
  • Page 44
    SPECIAL FUNCTIONS Chapter 3 Recommendation Ensure first Perfusor® Space pump is infusing Ensure first Perfusor® Space pump must be running in ‘continuous mode’ (i.e. ml/h or a dose rate; not KVO, PCA etc.) Deactivate Data Lock Data connection must be active between pumps –…
  • Page 45
    SPECIAL FUNCTIONS Chapter 3 Changes in TOM system: Change Reaction Rate changed in pump No user interaction necessary, will start infusion at new rate when syringe is empty. pump is stopped pump shows “connection lost – TOM aborted” alarm. TOM may be reactivated by pump is put in standby pressing and then…
  • Page 46: Chapter 4 Autoprogramming

    AUTOPROGRAMMING Chapter 4 A U TO P R O G R A M M I N G Note: All normal pump functions remain in place when orders are received via autoprogramming. The pump can accept drug orders wirelessly from the EHR system or from SpaceStation with SpaceCom.

  • Page 47
    AUTOPROGRAMMING Chapter 4 Note: Order may be cancelled prior to confirming order. • Once all values are confirmed, the Main Menu is displayed. Note: Soft Limit alert will be issued if value exceeds any soft limits set in drug library, soft limit may be overridden or value re-programmed per institutional policy.
  • Page 48
    AUTOPROGRAMMNG Chapter 4 • Follow prompt, pressing to accept order or key to cancel and hold order for later. New Primary Infusion: • To accept a new PRIMary order, stop infusion and clear current PRIMary infusion by pressing key and responding “yes” to clear current infusion. PIGGYback Orders: Orders received after PRIMary has been set will be for PIGGYback infusions, follow prompts on screen to stop the PRIMary to accept the PIGGYback order.
  • Page 49
    AUTOPROGRAMMING Chapter 4 Note: Changing values on any incoming order may only be done after confir- ming all values. Once all values are confirmed you may scroll to any value and open editor with to change value. Alternately, order may be cancelled and request made for revised order to be sent.
  • Page 50: Chapter 5 Options

    OPTIONS Chapter 5 OPTIONS The options functions may be selected and changed while the pump is infusing or stopped. To edit a menu item, select “Options” in the Main Menu and press l. Then select desired function with and follow the Instructions for Use as described. 5.1 Occlusion Pressure The higher the pressure level is set at, the higher the pressure level must rise before triggering an occlusion pressure alarm.

  • Page 51
    OPTIONS Chapter 5 OccluGuard activation / deactivation from the Main Menu • Go to Options Menu and press l. • Navigate through the list with and select OccluGuard. • OccluGuard can be activated with and deactivated with d. Pressure Leap/Drop detection The pressure leap/drop software detects sudden increases and decreases in infu- sion pressure respectively which can be caused by problems in IV access, or changes in pump position in the SpaceStation.
  • Page 52
    OPTIONS Chapter 5 OccluGuard Meaning Recommendation Symbol OccluGuard is active. Infusion is running stably Pending – OccluGuard has not enough data OccluGuard will automatically reactivate as soon as infusion rate OccluGuard Inactive drops below threshold levels – see above. Confirm alarm and check IV access, IV setup and syringe for cause of Occlusion has been occlusion.
  • Page 53: Data Lock

    OPTIONS Chapter 5 When a change is made to the infusion system (e.g. addition or removal of a pump to a SpaceStation, a change of infusion rate, a bolus application) the OccluGuard and pressure leap/drop are temporarily set to ‘pending’ ( ) to allow the system to reach a hydrostatic balance, and so prevent false alarms.

  • Page 54: Bolus Rate

    OPTIONS Chapter 5 Activation of the function: • Open data lock in Options Menu with l. • Select between level 1, 2 or 3 (if activated) with and confirm with k. • Enter code with and press in order to activate data lock. Changes to the protected values and the bolus function which are marked withy are only possible after entering the code.

  • Page 55: Alarm Volume

    OPTIONS Chapter 5 5.7 Alarm Volume Chose between 9 different alarm volume levels. • Open alarm volume in Options Menu with l. • Set volume with and confirm entry with k. 5.8 Date / Time • Open date/time in Options Menu with l. •…

  • Page 56: Chapter6 Alarms

    ALARMS Chapter 5 ALARMS The Perfusor® Space is equipped with a audible and optical alarm signal. Audible Alarm- Optical signal Staff call User confirmation type signal Red LED Yellow LED Text Device flashes device Press and follow Alarm alarm and the instruction on the alarm code display.

  • Page 57
    ALARMS Chapter 6 Display message Pre-alarm reason “Syringe nearly empty“ Very little fluid is left in syringe. “VTBI near end“ The preselected volume is nearly infused. “Time near end“ The preselected time is almost over. “Battery nearly empty“ The battery is almost discharged. “KVO mode“…
  • Page 58
    ALARMS Chapter 6 the alarm message, and the staff call (optional) are all cleared by pressing k. Corrections should be made in accordance with the alarm reason. Display message Alarm reason “Syringe empty“ There is no fluid left in the syringe. Due to varying syringe tolerances of syringes from other manufacturers, some fluid may be left inside the syringe.
  • Page 59
    ALARMS Chapter 6 “Calibrate device“ Pump calibration data have changed (e.g. after an update). Recalibrate device via the service program. “Claw malfunction“ The emergency release button was pressed and the claws manually opened. Take out syringe and contact technical service department. “Plunger plate not prop.
  • Page 60: Reminder Alarms

    ALARMS Chapter 6 6.3 Reminder Alarms Reminder alarms only occur in two cases: 1. A syringe is inserted, the pump doesn’t administrate, no value is being edited and the device is not operated for two minutes. An acoustic tone sounds, the yellow LED is constantly on and a staff call is activated.

  • Page 61: Battery Operation And Maintenance

    BATTERY OPERATION AND MAINTENANCE Chapter 7 BATTERY OPERATION AND MAINTENANCE The battery has an operating lifetime of 8 hours at 25 ml/h when new. For optimal treatment of the battery, the device is equipped with protection against overcharge and deep depletion. The battery pack is charged by the pump during connection to mains.

  • Page 62
    BATTERY OPERATION AND MAINTENANCE Chapter 7 • If a battery, which is not completely discharged, is charged several times, its capacity can be reduced. • Under normal temperature conditions a battery can be charged and discharged approx. 500 times before its lifetime decreases. •…
  • Page 63: Chapter 8 Compatible Syringes

    ) to ensure specific syringe brand compatibility. The Time to Occlusion alarm has been measured at 5 ml/h. The measured data are typical values which may vary because of possible syringe tolerances. Manufacturer: B. Braun Syringe Type Omnifix Omnifix Omnifix Omnifix Omnifix B.

  • Page 64
    COMPATIBLE SYRINGES Chapter 8 Manufacturer: TYCO USA Syringe Type Monoject Monoject Monoject Monoject Monoject Monoject 3 ml 6 ml 12 ml 20 ml 35 ml 50/60 ml TYCO USA Mat. No. 8881- 8881- 8881- 8881- 8881- 8881- 513934 516937 512878 520657 535762 560125…
  • Page 65
    COMPATIBLE SYRINGES Chapter 8 Manufacturer: TERUMO Syringe Type 3 ml 5 ml 10 ml 20 ml 30 ml 50 ml 60 ml TERUMO EU/USA/JAP Mat. No. 3SS*03L 3SS*05L 3SS*10L 3SS*20L 1SS*30LZ1 2BS-50LG 3SS*60L 1SS*05LZ1 1SS*10LZ1 SS*20ES typ. typ. typ. typ. typ.
  • Page 66
    COMPATIBLE SYRINGES Chapter 8 Manufacturer: Becton-Dickinson Syringe Type BD Precise BD Precise B-D Precise 50 ml A/P 20 ml A/P Mat. No. 300144 300141 Time to Occl. [mm:ss] 03:17 01:11 [mm:ss] 16:36 05:03 Manufacturer: Polfa Syringe Type Polfa 50 ml Mat.
  • Page 67
    COMPATIBLE SYRINGES Chapter 8 Syringes not specified in IEC/EN 60601-2-24 Nutrition pumps, in contrast to infusion pumps, are not classified as Class IIa according to the infusion pump norm IEC/EN 60601-2-24. There are therefore no direct guide- lines concerning the technical characteristics (accuracy of infusion rate, alarm parameters etc) of the relevant disposables.
  • Page 68: Start Up Graphs And Trumpet Curves

    START UP GRAPHS AND TRUMPET CURVES Chapter 9 START UP GRAPHS AND TRUMPET CURVES Start Up Curves Trumpet Curves The graphs show the accuracy/uniformity of flow in relation to time. They allow for the following: The delivery behaviour or delivery precision is essentially influenced by the type of (disposable syringe) used.

  • Page 69: Is Depending On Chapter10 Technical Data

    Moisture protection IP 22 (fluid protected for horizontal usage) External power supply: • Rated voltage Via B. Braun SpaceStation or optional mains adaptor (rated voltage 100 … 240 V AC~, 50/60 Hz) for stand alone operation • External low voltage 11 ……

  • Page 70
    Delivery rate 1 ml/h: KVO-rate = set rate (default setting 0.1 ml/h) Computer connection USB connection in combination with B. Braun interface lead CAN SP (8713230) including electrical insulation. Please pay attention to safety notices. History protocol < 3000 last history entries.
  • Page 71
    TECHNICAL DATA Chapter 10 • Only use combined with approved devices/accessories by the manufacturer, otherwise this may lead to higher emission or reduced immunity. • Use only compatible combinations of equipment, accessories, working parts and disposables with luer lock connectors. Essential Performance for Infusion pumps: •…
  • Page 72
    TECHNICAL DATA Chapter 10 EMC (ELECTROMAGNETIC COMPATIBILITY)
  • Page 73
    TECHNICAL DATA Chapter 10…
  • Page 74
    TECHNICAL DATA Chapter 10…
  • Page 75
    TECHNICAL DATA Chapter 10…
  • Page 76: Warranty / Tsc* / Service / Training / Cleaning / Disposal

    • the Technical Safety Checks are carried out regularly. Warranty The CE mark confirms that this B. Braun provides 24 months warranty, as from the date of delivery, for every medical product Perfusor® Space (12 months for every Battery-Pack SP). This covers repair or complies with the replacement of parts damaged as a result of design/manufacturing errors or «Council Directive…

  • Page 77
    Note: Do not use Hexaquart® or other alkylamine containing disinfectants. Recommended: disinfectant for wiping available from B. Braun: Meliseptol® Foam pure, Melsitt 10% and Melsept SF 10%. Note: Keep instrument upright and do not allow any part of instrument to become saturated with or submersed in fluid during cleaning operation.
  • Page 78
    Disposal The pumps as well as battery packs can be returned to B. Braun for further disposal. When taking care of disposing of disposables as well as infusion soluti- ons, please consider the applicable hygiene and disposal regulations.
  • Page 79: Instructions For Use Accessory

    Comfort additionaly includes a central alarm management and alarm LEDs. PoleClamp SP (8713130) A maximum of three B. Braun Space pumps and one SpaceControl can be stacked together when used with the PoleClamp SP. For detailed instructions on secure fixation of the PoleClamp SP please refer to «Overview Perfusor® Space»…

  • Page 80
    INSTRUCTIONS FOR USE ACCESSORY Chapter 12 3.) Push plug of Connection Lead SP into 12 V connector. Note: A maximum of three plugs can be stacked upon each other in socket P2. Battery-Pack SP (NiMH) (8713180) Battery-Pack SP (NiMH) incl. Pin (8713180A) For further information on the Battery-Pack SP (NiMH) see “Battery Operation”.
  • Page 81
    INSTRUCTIONS FOR USE ACCESSORY Chapter 12 Connection Lead SP (12 V) (8713231) Install the Connection Lead SP (12 V) in the following way: 1.) Connect plug to socket P2 on back of pump or F3 on SpaceStation respectively. 2.) Put the connection lead into the car socket. 3.) If necessary, remove red adaptor of motor vehicle connector by slightly turning and simultanously pulling.
  • Page 82
    INSTRUCTIONS FOR USE ACCESSORY Chapter 12 Caution: The user should always closely observe the local pump alarms as well. Note: A maximum of three plugs can be stacked upon each other in socket P2. Technical Data Connecting Wire white and green white and brown Alarm disconnected…
  • Page 83
    INSTRUCTIONS FOR USE ACCESSORY Chapter 12 P C A — A c c e s s o r i e s • Space PCA-Kit (REF 8713554) consisting of: : — Demand button — Hook and loop tape for fixation of the demand button at the patient s arm — Line fixation connection between hook and loop tape…
  • Page 84: Ordering

    ORDERING Art. No. B. Braun Perfusor® Space (100 – 240 V) ……….871 3030 Recommended accessories for the B. Braun Perfusor® Space: SpaceStation ………………..871 3140 SpaceCover Standard ………………871 3147 SpaceCover Comfort ………………871 3145 PoleClamp SP…………………871 3130 Power Supply SP EU III ……………….871 3110D Power Supply SP EU III 3.0m……………..871 3123D…

  • Page 85
    ORDERING 50ml, yellow inked cylinder and aspiration needle…….87 28801 F 50ml, yellow inked cylinder, aspiration needle and 15 µm particle filter …………….872 8800 F 50ml, black, aspiration needle and particle filter……..872 8828 F Omnifix® syringes Omnifix® 50/60 ml Luer Lock……………461 7509F Omnifix®…
  • Page 86
    ORDERING Type PCA, PVC tube 1,5 mm, 150 cm, Luer Lock ……..872 6019 with 0.2 µm Sterifix filter, PVC tube 1,5 mm, 200cm, Luer Lock …………………872 3001 with SafeSite valve, PVC tube 1,5 mm, 150 cm, Luer Lock………………..872 2820…
  • Page 90
    Manufactured by: B. Braun Melsungen AG B. Braun Melsungen AG Sparte Hospital Care 34209 Melsungen 34209 Melsungen Germany Germany Tel +49 (0) 56 61 71-0 Tel.: +49 (0) 56 61 71-0 Fax: +49 (0) 56 61 71-20 44 38916517 • Drawing No. I0688700201 Printed on pulp bleached 100 % chlorine-free www.bbraun.com…

Возможно, вам также будет интересно:

  • B braun perfusor compact s ошибка 022
  • B 2014 ошибка инфинити
  • B 1604 ошибка митсубиси
  • B 1570 ошибка рено
  • B 1309 ошибка jcb

  • Понравилась статья? Поделить с друзьями:
    0 0 голоса
    Рейтинг статьи
    Подписаться
    Уведомить о
    guest

    0 комментариев
    Старые
    Новые Популярные
    Межтекстовые Отзывы
    Посмотреть все комментарии